Determination of Selenium in Blood Serum by Inductively Coupled Plasma Atomic Emission Spectrometry with Pneumatic Nebulization

Varování

Publikace nespadá pod Filozofickou fakultu, ale pod Přírodovědeckou fakultu. Oficiální stránka publikace je na webu muni.cz.
Autoři

MACHÁT Jiří KANICKÝ Viktor OTRUBA Vítězslav

Rok publikování 2002
Druh Článek v odborném periodiku
Časopis / Zdroj Analytical and Bioanalytical Chemistry
Fakulta / Pracoviště MU

Přírodovědecká fakulta

Citace
Obor Analytická chemie, separace
Klíčová slova Selenium speciation; Spectrometry; ICP-AES; Pneumatic nebulization
Popis The possibility of determination of selenium in blood serum using inductively coupled plasma emission spectrometry with conventional pneumatic nebulization was studied. A high-resolution spectrometer (SBW = 6 pm) with laterally viewed ICP was employed. Analysis with conventional pneumatic nebulization could overcome laborious and demanding digestion, which is necessary for hydride generation. A pressure digestion with nitric acid at 160oC was sufficient to decrease the carbon content in serum sample to 5-10% of its original value. Spectral interference of the CN band was observed and mathematically corrected. It was found that the carbon-induced selenium line emission enhancement occurred even under ICP optimized conditions. A method of determination was developed and applied to the analysis of blood serum. True limit of detection in real samples is 0.01-0.02 mg/L and the limit of quantification (RSD 10%) is 0.03-0.07 mg/L using Se I 196.090 nm line at an integration time 10-2 s. The method was tested by analysis of porcine blood serum and serum reference material Seronorm MI 0181.
Související projekty:

Používáte starou verzi internetového prohlížeče. Doporučujeme aktualizovat Váš prohlížeč na nejnovější verzi.