Simultaneous System of DP-LIBS and Laser Ablation ICP-OES

Investor logo

Warning

This publication doesn't include Faculty of Arts. It includes Central European Institute of Technology. Official publication website can be found on muni.cz.
Authors

NOVOTNÝ Karel HRDLIČKA Aleš ŠTĚPÁNKOVÁ Kateřina KANICKÝ Viktor KAISER Jozef MALINA Radomír HAHN David

Year of publication 2012
Type Conference abstract
MU Faculty or unit

Central European Institute of Technology

Citation
Description The simultaneous DP-LIBS and Laser-Ablation ICP-OES setup was designed by utilizing a modified commercially available laser ablation system UP-266 Macro (New Wave). Specially designed ablation chamber was developed to allow combination LIBS, DP-LIBS and LA-ICP-OES technique. In preliminary results washout efficiency of new ablation chamber was verified. Basic signal comparison of DP-LIBS and double pulse LA-ICP-OES were studied on the urinary stone cross-section. Results demonstrate high sensitivity for both methods and high spatial resolution.
Related projects:

You are running an old browser version. We recommend updating your browser to its latest version.